MRU SWG200 CEM – Stationary Gas Analyzer
Spesifications :
O2 | (electrochemical) 0 … 25 % |
O2 | (paramagnetic) 0 … 25 % |
NO | (NDIR) 0 … 200/4.000 ppm |
NO2 | (NDIR) 0 … 150/1.000 ppm |
SO2 | (NDIR) 0 … 150/4.000 ppm |
CO2 | (NDIR) 0 … 50 % |
CO | (NDIR) 0 … 200/10.000 ppm |
N2O | (NDIR) 0 … 100/500 ppm |
CH4 | (NDIR) 0 … 500/10.000 ppm |
C3H8 | (NDIR) 0 … 200/5.000 ppm |
Zero point deviation | negligible due to automatic zeroing |
Deviation | less than 0.2% of the measuring range per month |
Calculated components | NOx: NO + NO₂, calculated ppm or mg / m3, user-selectable O₂ reference combustion calculations (efficiency, heat loss) on special request |
Display | Illuminated 3.5 “TFT color display |
Interfaces | Illuminated keyboard, password-protected operation 4 analog outputs 4 … 20 mA, galvanically isolated, max. load: 500 R 2 alarm relays, potential-free contacts: 24 Vdc, 5 A Data storage and data logger on SD card RS 485 digital interface (Modbus RTU DIN rail RS 485, to ProfiBus converter or to Ethernet converter |
Gas sampling probe | HD gas sampling probe, heated ceramic filter with backpurge, or gas sampling probe HD-GW, heated glass wool filter, or LD gas sampling probe, unheated with in-situ sintered metal filter, heated or unheated gas sampling line, PTFE DN 4/6 mm |
Gas conditioning | Thermoelectric gas cooler (Peltier) with constant +4 ° C dew point Teflon particle filter, internal Viton tubing Monitored and regulated gas sampling pump Constant gas flow of 50 l/h Gas inlet pressure: -200 … +20 mbar (hPa) Sample gas outlet: atmospheric pressure |
Housing | Stainless steel housing, continuously monitored housing ventilation with alarm, antifreeze heater 200 W (option) |
Operating temperature | (+5 … +45 °C) or (-10 … +45 °C) with housing heating |
Power supply | Universal: 90 … 240 Vac, 47 … 63 Hz, 120 W (420 W with heating) |
Protection | IP 54 |
Dimensions | (W x H x D) 600 x 700 x 210 mm, suitable for wall mounting |
Weight | approx. 50 kg |
SWG200 – Continous Emission Monitoring System
The MRU SWG 200 CEM is a high-performance stationary gas analyzer for continuous flue gas and emission monitoring in industrial applications such as engines, boilers, and waste-to-energy plants.
Key Features :
-
Simultaneous NDIR measurement of up to 8 gases, including NOx, CO, CO₂, SO₂, CH₄, N₂O, C₃H₈, and O₂.
-
Uses advanced NDIR technology with excellent accuracy and zero-drift performance.
-
O₂ measured via electrochemical or paramagnetic sensor.
-
Includes automatic calibration, zeroing, and a double-stage Peltier gas cooler with dual condensate pumps.
-
Optimized for cold/dry gas sampling with low sample flow (~1 L/min).
-
Stainless steel cabinet, IP54-rated, with indoor/outdoor installation options.
-
Integrated data interfaces: RS485 (Modbus RTU), Profibus, Ethernet, USB, and SD card.
Applications :
- Natural gas/diesel engines
- Biomass and bagasse boilers
- Biogas/landfill gas CHP
- Cement outlet stacks
- Coal power stations, etc
In summary, the SWG 200 CEM is a robust and versatile analyzer ideal for demanding emission monitoring tasks, offering precise multi-gas detection and advanced data handling.