MRU SWG200 CEM – Stationary Gas Analyzer

Spesifications :

O2 (electrochemical) 0 … 25 %
O2 (paramagnetic) 0 … 25 %
NO (NDIR) 0 … 200/4.000 ppm
NO2 (NDIR) 0 … 150/1.000 ppm
SO2 (NDIR) 0 … 150/4.000 ppm
CO2 (NDIR) 0 … 50 %
CO (NDIR) 0 … 200/10.000 ppm
N2O (NDIR) 0 … 100/500 ppm
CH4 (NDIR) 0 … 500/10.000 ppm
C3H8 (NDIR) 0 … 200/5.000 ppm
Zero point deviation negligible due to automatic zeroing
Deviation less than 0.2% of the measuring range per month
Calculated components NOx: NO + NO₂, calculated ppm or mg / m3, user-selectable O₂ reference
combustion calculations (efficiency, heat loss) on special request
Display Illuminated 3.5 “TFT color display
Interfaces Illuminated keyboard, password-protected operation
4 analog outputs 4 … 20 mA, galvanically isolated, max. load: 500 R
2 alarm relays, potential-free contacts: 24 Vdc, 5 A
Data storage and data logger on SD card
RS 485 digital interface (Modbus RTU
DIN rail RS 485, to ProfiBus converter or to Ethernet converter
Gas sampling probe HD gas sampling probe, heated ceramic filter with backpurge,
or gas sampling probe HD-GW, heated glass wool filter,
or LD gas sampling probe, unheated with in-situ sintered metal filter,
heated or unheated gas sampling line, PTFE DN 4/6 mm
Gas conditioning Thermoelectric gas cooler (Peltier) with constant +4 ° C dew point
Teflon particle filter, internal Viton tubing
Monitored and regulated gas sampling pump
Constant gas flow of 50 l/h
Gas inlet pressure: -200 … +20 mbar (hPa)
Sample gas outlet: atmospheric pressure
Housing Stainless steel housing, continuously monitored housing ventilation with alarm, antifreeze heater 200 W (option)
Operating temperature (+5 … +45 °C) or (-10 … +45 °C) with housing heating
Power supply Universal: 90 … 240 Vac, 47 … 63 Hz, 120 W (420 W with heating)
Protection IP 54
Dimensions (W x H x D) 600 x 700 x 210 mm, suitable for wall mounting
Weight approx. 50 kg

Download Brochure

Description

SWG200 – Continous Emission Monitoring System

The MRU SWG 200 CEM is a high-performance stationary gas analyzer for continuous flue gas and emission monitoring in industrial applications such as engines, boilers, and waste-to-energy plants.

Key Features :

  • Simultaneous NDIR measurement of up to 8 gases, including NOx, CO, CO₂, SO₂, CH₄, N₂O, C₃H₈, and O₂.

  • Uses advanced NDIR technology with excellent accuracy and zero-drift performance.

  • O₂ measured via electrochemical or paramagnetic sensor.

  • Includes automatic calibration, zeroing, and a double-stage Peltier gas cooler with dual condensate pumps.

  • Optimized for cold/dry gas sampling with low sample flow (~1 L/min).

  • Stainless steel cabinet, IP54-rated, with indoor/outdoor installation options.

  • Integrated data interfaces: RS485 (Modbus RTU), Profibus, Ethernet, USB, and SD card.

Applications :

  • Natural gas/diesel engines
  • Biomass and bagasse boilers
  • Biogas/landfill gas CHP
  • Cement outlet stacks
  • Coal power stations, etc

In summary, the SWG 200 CEM is a robust and versatile analyzer ideal for demanding emission monitoring tasks, offering precise multi-gas detection and advanced data handling.

About brand
MRU GmbH, founded in 1984 in Germany, is a globally recognized manufacturer specializing in the development and production of gas analysis and environmental monitoring instruments. Renowned for its precision-engineered portable and stationary gas analyzers, MRU serves a wide range of industries including power plants, cement production, waste incineration, biogas, HVAC, and environmental agencies. MRU’s instruments are designed to measure gases such as O₂, CO, NOx, SO₂, CH₄, CO₂, H₂S, and also support dust and flow monitoring, all while complying with leading international standards like EN, ISO, and US EPA. The company is known for its commitment to German-quality engineering, long-term reliability, and user-friendly design, providing trusted solutions for combustion control, emissions monitoring, and air quality management.